AnBei is a 60 GHz mmWave breathing and heart-rate recorder for children's rooms. The sensor measures a child's breathing and heart-beat motion without contact. But almost every disappointing field result I have seen came down to one of two things: the geometry was wrong, or something was in the way. This post is the mounting checklist we ended up with — the parts that are physics, not marketing.

The measurement is tiny

At 60 GHz the wavelength is about 5 mm. Breathing moves the chest wall 4–12 mm; the heartbeat component is roughly 0.1–0.5 mm. The radar does not measure those displacements directly. It measures phase shift in the returned signal, and the phase is much more sensitive than the range resolution — which is why sub-millimetre motion is detectable at all.

That sensitivity is also the problem. Anything that changes the RF path looks like motion.

Geometry: a cone, not a room

The usable volume is a cone, not "wherever the sensor can see".

  • Height: 1.4–1.6 m above the floor, on the wall at the head of the bed.
  • Angle: about 45° downward, aimed at the child's chest, not the mattress centre.
  • Slant range: ≤ 1.5 m from the sensor to the chest surface.

Two failure modes dominate:

  1. Too low. A sensor lying on a bedside table looks horizontally across the room. The chest wall moves toward and away from the sensor along the line of sight — from the side, the radial component collapses and the phase signal with it. People report "it sees presence but not breathing". That is exactly what a horizontal placement produces.
  2. Too far. Phase noise grows with range. At 2.5–3 m the heartbeat component is buried; you may still get a breathing rate, but it will be unstable.

Occlusion: RF is not magic

Every one of these attenuates or scatters the signal enough to break the measurement:

  • bed rails and side rails
  • mosquito nets and bed canopies
  • thick curtains or a wardrobe door left open in the path
  • a wall between sensor and bed (60 GHz does not penetrate drywall usefully)
  • a ceiling fan or air conditioner blowing across the path

The practical rule: if you can draw a straight line from the sensor face to the child's chest and that line crosses anything, expect degraded or missing readings. This is also why "just put it behind the headboard" does not work.

When a reading is invalid

The device reports a value; that does not mean the value means anything. Readings are only meaningful when the child is:

  • lying in bed,
  • not moving (rolling over invalidates the current sample),
  • alone in the bed,
  • within the slant-range limit.

Sitting up, walking around, being held, or sharing the bed with a parent all produce numbers that should be discarded rather than averaged. A recording product should mark those windows as invalid instead of smoothing them into the nightly curve.

The offset you have to be honest about

Our heart-beat estimate runs about 18 beats per minute high in absolute terms. That is a known bias of the phase-based extraction at this range, not a calibration we can fix by multiplying by a constant — the bias is not stable across subjects and postures.

So the product only claims trends: this week versus last week, for the same bed and the same child. Absolute BPM is shown with that caveat, and we do not use it to classify anything.

Breathing rate is tighter — within about ±2 breaths per minute under the conditions above.

A mounting checklist

Before trusting a single night of data:

  1. Sensor at 1.4–1.6 m, 45° down, aimed at the chest.
  2. Slant range measured, not estimated — ≤ 1.5 m.
  3. Nothing in the line of sight: no net, no canopy, no rail, no glass, no wall.
  4. One person in the bed, lying still during the calibration window.
  5. Record a baseline night, then compare night-to-night on the same setup.
  6. Treat any night with a changed layout as a new baseline.

Boundaries

AnBei is a recording device, not a medical device. It does not diagnose, does not detect apnoea, does not measure blood oxygen, and is not a substitute for supervision. Children aged 0–3 must always have a caregiver in the same room. If a child shows any abnormality, seek medical attention.

The engineering point is narrower than the marketing usually is: a 60 GHz radar in a child's room is a geometry problem first and a signal-processing problem second. Get the cone right and the occlusion wrong, and you get nothing. Get both right, and the phase data is genuinely usable for overnight trends.